RS232_485_ETP_MODBUS_BU_REV02.doc
Set command: DLTTI=
n
0:OK if parameters accepted
1:CMD ERR if logging functions not
enabled
2:PARAM ERR if
n
out of range
5:ACCESS ERR if insufficient access
level
Help command: DLTTI=?
min <> max (units)
if accepted
1:CMD ERR if logging functions not
enabled
5:ACCESS ERR if insufficient access
level
Read command: DTSON?
Status of operation if accepted
5:ACCESS ERR if insufficient access
level
Set command: DTSON=1
0:OK if parameters accepted
2:PARAM ERR if
n
not equal to 1
3:EXEC ERR if operation not possible
5:ACCESS ERR if insufficient access
level
DTSON (Data logger Test Start
ON) Starts the “step check” test
mode and reads the status of the
operation (1 = in progress , 0 =
terminated). If the stop date is less
than the current time and date, an
error code is returned
Help command: DTSON=?
1:EXECUTE
if accepted
5:ACCESS ERR if insufficient access
level
Read command: NOT
SUPPORTED
1:CMD ERR if command is issued
Set command: DTSOF=1
0:OK if parameters accepted
2:PARAM ERR if
n
not equal to 1
5:ACCESS ERR if insufficient access
level
DTSOF (Data logger Test Start
OFf) Stops or suspend the “step
check” test mode.
Help command: DTSOF=?
1:EXECUTE
if accepted
5:ACCESS ERR if insufficient access
level
COMMANDS AND FUNCTIONS EQUIVALENT TO THE “10-DIAGNOSTIC” MENU
Name and description
Modes
Returned values or codes
Read command: CALIC?
Status of calibration operation if
accepted
Set command: CALIC=1
0:OK if parameters accepted
2:PARAM ERR if
n
not equal to 1
CALIC (CALibration Immediate
Command) performs an immediate
calibration cycle and reads the
status of the operation (1 = in
progress , 0 = terminated).
Help command: CALIC=?
1:EXECUTE
if accepted
Read command: NOT
SUPPORTED
1:CMD ERR if command is issued
Set command: ATSIC=1
0:OK if parameters accepted
2:PARAM ERR if
n
not equal to 1
5:ACCESS ERR if insufficient access
level
ATSIC
(Auto-TeSt
Immediate
Command) Performs an auto-test
cycle.
Help command: ATSIC=?
1:EXECUTE
if accepted
5:ACCESS ERR if insufficient access
level
MSIEN
(Measure
SImulation
ENable) Enables or disables the
measure simulation function.
Read command: MSIEN?
Status of measure sim. function if
accepted
5:ACCESS ERR if insufficient access
level