Models 5800-s – Brooks Instrument 5800S Series User Manual

Page 26

background image

Installation and Operation Manual

X-TMF-5800S-MFC-eng

PN 541-C-051-AAG

November, 2008

Models 5800-S

26

Temperature

Both ambient and process gas: 0-70 °C.

Leak Integrity

Outboard: 1 x 10

-9

mbar l/sec.

Helium

Warm-up time

< 10 minutes; 1% F.S. accuracy.
Performance within specifications: 45 minutes.

Damping

Damping from 0 to 10 seconds is possible for the flow output
signal(s) * (default setting is 0.5 sec.).

Response

Standard response of the flow output signal 1 sec.

(Meters)

Response up to 0.2 sec. on request.
Model 5864 3sec. (1 sec. on request)

Settling Time

Standard 1 sec. Settling time or up to 0.2 seconds is

(Controllers)

possible (to within 2% full scale of final value) for any
command (Set point) step; virtually without any dead time,
over- or undershoot. Model 5853S 3 sec. (1 sec. on
request).

* To be specified at ordering

Section 5: Specification