6.3.9 SYSTem Subsystem

SYSTem:ERRor Command

Command syntax:

:SYSTem:ERRor?

Description:

This query command is used to query error messages from the error queue. From front display, if

a wrong command has been sent, the display will show an error message for approximately one

second. That error can be queried from the error queue by sending this command.

Error Messages:
“NO ERROR!” – This means no error(s) have occurred.
“BUS:BAD COMMAND.” – This means either a wrong command was sent to the instrument or there may
be misspelling in the command.

6.3.10 Common Commands

Common commands can be used to all equipment. We provide some common commands as shown

below:

*RST

Command Syntax:

*RST

Description:

Reset the instrument

*TRG

Command Syntax:

*TRG

Description:

Trigger the instrument for measurement

*IDN?

Query Syntax:

*IDN?

Query return:

,

Here:

2831E Multimeter

Ver1.0.09.12.03

Description: Query information of the instrument