1 beeper control, 2 baud rate – B&K Precision 5491B - Manual User Manual
Page 38

38
Terminal character setup. The information is not directly related to making measurements but is an
important part of operating the multimeter.
4.5.1 Beeper Control
Normally, the multimeter will emit a tone whenever certain conditions are met. For example: the
multimeter will beep when a stable reading is captured in reading hold. You may want to disable the
beeper for certain applications.
When you disable the beeper, the multimeter will not emit a tone when:
1. A limit is exceeded in a limit test
2. A stable reading is captured in reading hold.
Disabling the beeper has no effect on the tone generated when:
1. An internal error is generated.
2. The continuity threshold is exceeded.
3. A front panel key is pressed.
The beeper state is stored in non-volatile memory and does not change when power has been off
or after a reset. The beeper is enabled when the multimeter is shipped from the factory.
Use the following steps to change the b
eeper’s state:
1. Press
→
to enter the menu on the menu level,
“A: MATH MEU” will be displayed.
2. Use
or
key to move across to the SYS MEU on the menu level,
“C: SYS MEU” will
be displayed.
3. Press
to move down to the command level within the SYS MEU.
4. Use
or
key to move across to the BEEP command,
“1: BEEP STA” will be
displayed.
5. Press
to move down a level to set the beeper control.
6. Use
or
key to select ON or OFF.
7. Press
(ENTER) to confirm the beeper control. The message
“SAVED” will be displayed to
show that the change is now in effect. The multimeter will exit the parameter level and move up a
level to the command level.
8. Press
→
key to exit from the menu and return to the measurement status.
4.5.2 Baud rate
The baud rate is the rate at which the digital multimeter and the computer communicate. Choose one of
these available rates:
38.4k
19.2k
9600
4800
2400