B&K Precision 5491B - Manual User Manual
Page 61
61
:RANGe
Command syntax:
:FREQuency:THReshold:VOLTage:RANGe
:PERiod:THReshold:VOLTage:RANGe
Set voltage threshold range.
Command parameter:
Specify signal levels in volts (voltage threshold)
Query:
:RANGe?
Query maximum signal level.
Description:
These commands are used to specify the expected input level. The instrument will then
automatically select the most ideal current or voltage threshold range.
:REFerence
:REFerence
Command syntax:
:FREQuency:REFerence
Specify reference for FREQ
:PERiod:REFerence
Specify reference for PER
Command parameter:
0 to 1.0e6 Reference for FREQ
0 to 1 Reference for PER
DEFault
0 (All measurement functions)
MINimum Minimum value for specified function
MAXimum Maximum value for specified function
Query:
:REFerence?
Query reference for relative function
Description:
These commands are used to establish a reference value for the specified function. When reference
is enabled (:REFerence:STATe), the result will be the algebraic difference between the input signal
and the reference value:
Reading = Input signal
– Reference
From the front panel, reference is called relative (REL).
The :REFerence
(:REFerence
the :REFerence
Conversely, when a reference is set using the :ACQuire command, the :REFerence? query
command returns the acquired reference value.