B&K Precision 5491B - Manual User Manual
Page 55
55
Command parameter:
= 1 or ON
Enable auto range
0 or OFF
Disable auto range
Query:
:AUTO?
Query auto range (ON or OFF)
Description:
These commands are used to control auto ranging. With auto ranging enabled, the instrument
automatically goes to the most ideal range to perform the measurement.
The auto range command (:RANGe:AUTO) is coupled to the command that manually selects the
measurement range (:RANGe
instrument remains at the automatically selected range. When a valid :RANGe
sent, auto ranging disables.
:REFerence
:REFerence
Command syntax:
: CURRent:AC:REFerence
Specify reference for ACI
: CURRent:DC:REFerence
Specify reference for DCI
Command parameter:
Reference for ACI
0 to 20
Reference for DCI
DEFault
0 (All measurement functions)
MINimum Minimum value for specified function
MAXimum Maximum value for specified function
Query:
:REFerence?
Query reference for relative function
Description:
These commands are used to establish a reference value for the specified function. When reference
is enabled (:REFerence:STATe), the result will be the algebraic difference between the input signal
and the reference value:
Reading = Input signal
– Reference
From the front panel, reference is called relative (REL).
The:REFerence
(:REFerence
the :REFerence