B&K Precision 5491B - Manual User Manual
Page 58

58
:RANGe Commands
:[UPPer]
Command syntax:
:RESistance:RANGe[:UPPer]
Set measurement range for
Ω
Command Parameter:
Expected reading is ohms
DEFault 20e6 (
Ω)
MINimum
0 (All functions)
MAXimum Same as DEFault
Query:
:RANGe[:UPPer]?
Query measurement range of the current function.
Description:
This command is used to manually select the measurement range for the specified measurement
function. The range is selected by specifying the expected reading as an absolute value. The
instrument will then go to the most ideal range that will accommodate the expected reading. For
example, if you expect a reading of approximately 20
Ω, simply let the parameter
to select the 200
Ω range (for model 2831E). If you expect a reading of approximately 50 Ω, simply
let the parameter
Ω range (for model 5491B).
:AUTO
Command syntax:
:RESistance:RANGe:AUTO
Set auto range for
Ω
Command parameter:
= 1 or ON
Enable auto range
0 or OFF
Disable auto range
Query:
:AUTO?
Query auto range (ON or OFF)
Description:
These commands are used to control auto ranging. With auto ranging enabled, the instrument
automatically goes to the most ideal range to perform the measurement.
The auto range command (:RANGe:AUTO) is coupled to the command that manually selects the
measurement range (:RANGe
instrument remains at the automatically selected range. When a valid :RANGe
sent, auto ranging disables.