24

Note:

Source current flows from the

to COM terminals

Figure 3-3

Resistance Measurements

3.5 Measuring Frequency and Period

Frequency measurement range: 5 Hz to more than 1 MHz.

Period measurement range: 0.2 s to less than 1

μs.

Input signal range: 200 mV AC to 750V AC.

The instrument uses the volts input terminals to measure frequency. The AC voltage range can be

changed with the RANGE

and

keys. However, the signal voltage must be greater than

10% of the full-scale range.

3.5.1 Trigger Level and Measurement Errors

The multimeter uses a technique which maintains a constant resolution for any input frequency to

measure frequency and period. The gate time is always a multiple of the measured signal period rather

than a fixed time. The error will be no more than +/-1 from the total gate counts, which assures an

equivalent accuracy over the whole frequency range.

3.5.2 Gate Time

Gate time is the amount of time the meter uses to sample frequency or period readings. The

measurement speed rate and the measuring frequency change the gate time.

3.5.3 Connections

Assuming the multimeter is under factory default conditions, the basic procedure for measuring

frequency and period is as follows:

1. Connect test leads to

and COM terminals.

2. Select frequency or period measurement functions by pressing

or

3. Connect test leads to the source as shown in Figure 3-4:

CAUTION: Do not exceed 1000 V peak between

and COM, or instrument may be

FREQ

FREQ