6 measuring continuity – B&K Precision 5491B - Manual User Manual
Page 25
25
damaged.
4. Take reading from the display
Figure 3-4 Frequency and Period Measurements
Input Impedance =1.1
MΩ in parallel with 100 pF
CAUTION: Maximum Input = 750 V RMS or 1000 V peak, 3x10
7
V-Hz
3.6 Measuring Continuity
Under continuity test, when 200
Ω range is selected, the multimeter alerts you with a beep when a
reading is below a threshold resistance level of 10
Ω. For other ranges, the beeper will sound if a reading
is below a threshold resistance level listed in table 3-1. This beep will sound within less than half a
second when reading measured is below the threshold.
Table 3-1 Threshold resistance levels in Continuity Test
Model 2831E
Measuring range
Beeper on
200.00
Ω
<10
Ω
2.0000 k
Ω
<100
Ω
20.000 k
Ω
<1
kΩ
200.00 k
Ω
<10
kΩ
2.0000
MΩ
<100
kΩ
20.000
MΩ
<1
MΩ
Model 5491B
Measuring range
Beeper on
500.00
Ω
<10
Ω
5.0000 k
Ω
<100
Ω
50.000 k
Ω
<1
kΩ
500.00 k
Ω
<10
kΩ
5.0000 M
Ω
<100
kΩ
50.000
MΩ
<1
MΩ
AC VOLTAGE SOURCE