:REFerence Commands

:REFerence

Command syntax:

:RESistance:REFerence

Specify reference for

Ω

Command parameter:

= 0 to 20e6

Reference for Ω

DEFault

0 (All measurement functions)

MINimum Minimum value for specified function

MAXimum Maximum value for specified function

Query:

:REFerence?

Query reference for relative function

Description:

These commands are used to establish a reference value for the specified function. When reference

is enabled (:REFerence:STATe), the result will be the algebraic difference between the input signal

and the reference value:

Reading = Input signal

– Reference

From the front panel, reference is called relative (REL).

The :REFerence command is coupled to the :ACQuire command. The last command sent

(:REFerence or :ACQuire) establishes the reference. When a reference is set using

the :REFerence command, the REFerence? query command returns the programmed value.

Conversely, when a reference is set using the :ACQuire command, the :REFerence? query

command returns the acquired reference value.

:STATe

Command syntax:

:RESistance:REFerence:STATe

Control reference for

Ω

Command parameter:

= 1 or ON

Enable reference

0 or OFF

Disable reference

Query:

:STATe?

Query state of reference.

Description:

These commands are used to enable or disable

“reference” for the specified function. When

enabled, the displayed reading will include the programmed reference value. When disabled, the

displayed reading will not include the reference value.