6.3.2 FUNCtion subsystem
The commands in this subsystem are used to configure the measurement function subsystems and are
summarized in Table 6-2.
Table 6-2
FUNCtion Subsystem Commands Summary
Commands
Selectable Functions
:FUNCtion
:FUNCtion?
Select measurement function: VOLTage:AC, VOLTage:DC,
RESistance, CURRent:AC, CURRent:DC, FREQuency, PERiod,
DIODe, CONTinuity.
Query Function
:FUNCtion Command
:FUNCtion
Command syntax:
:FUNCtion
Command Parameter:
VOLTage:AC
Select AC Voltage
VOLTage:DC Select DC Voltage
CURRent:AC Select AC Current
CURRent:DC
Select DC Current
RESistance Select 2-wire Resistance
FREQuency Select Frequency
PERiod Select Period
DIODe Select Diode Testing
CONTinuity
Select Continuity Testing
Query:
:FUNCtion?
Query currently selected function
Description:
This command is used to select the measurement function of the instrument.
Each measurement function “remembers” its own unique setup configuration, such as range, speed,