B&K Precision 5491B - Manual User Manual
Page 51

51
= 1 or ON
Enable auto range
0 or OFF
Disable auto range
Query:
:AUTO?
Query auto range (ON or OFF)
Description:
These commands are used to control auto ranging. With auto ranging enabled, the instrument
automatically goes to the most ideal range to perform the measurement.
The auto range command (:RANGe:AUTO) is coupled to the command that manually selects the
measurement range (:RANGe
instrument remains at the automatically selected range. When a valid :RANGe
sent, auto ranging disables.
:REFerence
:REFerence
Command syntax:
:VOLTage:AC:REFerence
Specify reference for ACV
:VOLTage:DC:REFerence
Specify reference for DCV
Command parameter:
Reference for ACV
-1010 to 1010 Reference for DCV
DEFault
0 (All measurement functions)
MINimum Minimum value for specified function
MAXimum Maximum value for specified function
Query:
:REFerence?
Query reference for relative function
Description:
These commands are used to establish a reference value for the specified function. When reference
is enabled (:REFerence:STATe), the result will be the algebraic difference between the input signal
and the reference value:
Reading = Input signal
– Reference
From the front panel, reference is called relative (REL).
The :REFerence
(:REFerence
the :REFerence
Conversely, when a reference is set using the :ACQuire command, the :REFerence? query
command returns the acquired reference value.
(Note: In order to send a REFerence? query, the reference value must be set either