beautypg.com

B&K Precision 5491B - Manual User Manual

Page 47

background image

47

Table 6-1

DISPlay Subsystem Commands Summary

Command

Function Description

:DISPlay

:ENABle

:ENABle?

Enable or disable front panel display

Query state of the display

:ENABle

Command syntax:

:DISPlay:ENABle

Command Parameter:

=

0 or OFF

Disable front panel display

1 or ON

Enable front panel display

Query:

:ENABle? Query state of the display

Description:

This command is used to enable or disable front panel display circuitry. When disabled, the

instrument operates at a higher speed. While disabled, the display is frozen. All front panel controls

except LOCAL are disabled. Normal display operation can be resumed by using:ENABle command

or pressing LOCAL key to enable the display.