M&C Products Analysentechnik GmbH
9-3.12-ME
- 34 -
18.Trouble shooting
Fault description
Display
Solution
Measuring cell doesn‘t work
E:01 resp. F:01, depending on
Give air resp. nitrogen to the analyser
properly at 100 Vol.-%
the language
and restart the analyser
(suppressed measuring
range 99-100 Vol.-%)
Calibration fault:span gas is
Display of the calibration fault
Start new calibration.
switched off while proceed-
until a new calibration is
ing the calibration of the
carried out; the displayed value
span.
is related to the last calibration.
Calibration limits of the A/D-
Check the limits (see below) testing the
converter exeeded.
converter channel (see menu);
When limits are exeeded a calibration at
the manufacturer has to be carried out.
Limits testing the converter channel (see menu):
Checking the zero point with nitrogen N
2
U
min
[V]
U
max
[V]
0,005
0,147
Checking the span with ambient air (20,93 Vol.-%) at
normal pressure and different voltages of the zero point:
Zero point [V]
U
min
[V]
U
max
[V]
0,005
0,306
0,344
0,078
0,370
0,398
0,147
0,430
0,457
Checking the span with 100 Vol.-% oxygen O
2
at normal
pressure and different voltages of the zero point:
Zero point [V]
U
min
[V]
U
max
[V]
0,005
1,440
1,580
0,078
1,470
1,610
0,147
1,500
1,630