18.Trouble shooting

Fault description

Display

Solution

Measuring cell doesn‘t work

E:01 resp. F:01, depending on

Give air resp. nitrogen to the analyser

properly at 100 Vol.-%

the language

and restart the analyser

(suppressed measuring

range 99-100 Vol.-%)

Calibration fault:span gas is

Display of the calibration fault

Start new calibration.

switched off while proceed-

until a new calibration is

ing the calibration of the

carried out; the displayed value

span.

is related to the last calibration.

Calibration limits of the A/D-

Check the limits (see below) testing the

converter exeeded.

converter channel (see menu);

When limits are exeeded a calibration at

the manufacturer has to be carried out.

Limits testing the converter channel (see menu):

Checking the zero point with nitrogen N

2

U

min

[V]

U

max

[V]

0,005

0,147

Checking the span with ambient air (20,93 Vol.-%) at
normal pressure and different voltages of the zero point:

Zero point [V]

U

min

[V]

U

max

[V]

0,005

0,306

0,344

0,078

0,370

0,398

0,147

0,430

0,457

Checking the span with 100 Vol.-% oxygen O

2

at normal

pressure and different voltages of the zero point:

Zero point [V]

U

min

[V]

U

max

[V]

0,005

1,440

1,580

0,078

1,470

1,610

0,147

1,500

1,630