9-3.12-ME

M&C Products Analysentechnik GmbH

- 17 -

↑↑

input config.

input 1...4

↑↑

average

no, 1, ..., 100 seconds

An alarm situation can be kept. If yes, one input must be
configured as a reset (see 12.4.9).
The entry starts with the Enter key. The first changeable
position is underlined and can be changed operating the
Direction keys

⇑⇓

. A jump to the next position happens via

the Select key.
The entry has to be confirmed by the Enter key, and the

Select key leads to the next menu point.

12.4.9 Input configuration
Inputs 1 to 4 are selected by the Direction keys

⇑⇓

(see 7.5).

The Enter key opens the entry. The Direction keys

⇑⇓

select

one of the following input functions:
• external alarm (passing through an incoming signal),
• set span 1 to 4,
• autocalibration offset,
• autocalibration,
• alarm reset,

resets all alarms, which are setted to ‘keep alarm=1’ in
the previous menu point. The signal for the reset function
must be an impulse.

The choice has to be confirmed via the Enter key.

The Select key leads to the next menu point.

12.4.10Average
This menu point opens the possibility to determine an
average value over a maximum measuring time of 100
seconds. This average value is displayed. The desired
value can be selected via the Direction keys

⇑⇓

.

The Enter key operates the confirmation and the Select
key leads to the next menu point.

Select

Enter

⇑⇑⇑⇑⇑

⇓⇓⇓⇓⇓

keep alarm = 1

1:0 2:0 3:0 4:0

Enter

⇑⇑⇑⇑⇑

⇓⇓⇓⇓⇓

input config.

external alarm

Enter

⇑⇑⇑⇑⇑

⇓⇓⇓⇓⇓

Select

Enter

⇑⇑⇑⇑⇑

⇓⇓⇓⇓⇓

Select

Enter

Select