M&C TechGroup PMA 100 Operator's manual User Manual
Page 26

M&C Products Analysentechnik GmbH
9-3.12-ME
- 26 -
cal. log-book
M 10.11.99
11:o7
Select
calib. diff.
ofs: (-)x.x%
g: x.x%
Select
inflate gas
. . . sec.
21.00
d: (-) 00.00
Afterwards the analyser is purged with sample gas.
After finishing the procedure the analyser jumps automatical-
ly into the measurement level.
A fault during the calibration is displayed as:
E
... O
2
-calibration, and is also stored in the calibration log-
book.
The next menu point can be reached stepping into the
calibration mode via the Cal key and passing the previous
menu points with the Select key.
12.8.4 Deviation from calibration
The display shows the deviation from the basic calibration
carried out in the factory. The offset is given of the possible
deviation (
±
100%).
Select leads to the following menu point.
12.8.5 Calibration log-book
It is possible to read back the data (date and time) of the
last 9 calibrations via the Direction keys
⇑⇓ .
The kind of calibration is marked as follows:
M manual calibration,
O manual calibration offset,
S autocalibration offset,
A autocalibration.
A fault during the calibration is displayed as (e.g.) :
E
O
10.11.9911:07
Select leads to the beginning of the calibration
Select
Cal