12.2
Warming up
After switching on the analyser type PMA100, the warming-
up begins.
The following display appears for approximately 20 seconds:
• M&C logo
• current O
2
-concentration
• Software-version implemented in the PMA100
• and the trademark.
After further 20 seconds the display changes. Via the Direc-
tion keys
⇑⇓
the following current values can be indicated:
• measuring range,
• temperatur of the transducer,
• pressure in the measuring-cell,
• date/time,
• current output.
The second line displays the heating procedure.
The warming-up is finished by reaching a temperature of
54
°
C. The display shows either the current value (see
above) or an error message. The second line of the display
presents the current oxygen concentration (for multi-channel
version the concentration of the respective channel).
The Select key leads to the next menu point.
12.3
Alarm log-book
The starting time and the 9 last alarms are stored. The
Direction keys
⇑⇓
can be used going through the alarm log-
book. The current alarm message is shown on the display
like
E
....
The Select key leads to the entry of the access code.
The Enter key opens the access code menu for the first,
second respectively read only level.
The first changeable position is underlined. With the Direc-
tion keys
⇑⇓
the value of the position can be changed in
between 0 and 9. A jump to the next position happens via
the Select key. The entry of the complete code has to be
confirmed by the Enter key and is shown in the second line
of the display.
The codes are preadjusted at the factory:
Code: 0010
level 1
Code: 1000
level 2
Code: 0000
read only-level;
The read only-level allows a passage through all menu
points of level 1 via the Select key. In this level parameters
can not be changed.
The Select key leads to level 1 - measuring parameters,
relay-, input and output configuration.
MC
Para O
2
V 1.52
TM M&C Products
Enter
enter to config.
Code: 0000
Select
enter to config.
Code: 0000
Select
Select
⇑⇑⇑⇑⇑
⇓⇓⇓⇓⇓
press:
1005 mbar
heating 22.5
°
C
⇑⇑⇑⇑⇑
⇓⇓⇓⇓⇓
Enter
enter to config.
Level 1
press:
1005 mbar
alarm log-book
Select
⇑⇑⇑⇑⇑
⇓⇓⇓⇓⇓