Example :
The residual oxygen percentage should be measured in a
closed 100% carbon dioxide (CO

2

) atmosphere. The „zero-

calibration“ is done by means of Nitrogen (N

2

).

According to the list of cross-sensitivities the error for CO

2

at 20

°

C is -0,27%. In order to obtain a higher accuracy this

means for the calibration that the reading should be
adjusted at +0,27% with N

2

, in order to compensate the

error of CO

2

.

Since the values of cross-sensitivities are based on 100 Vol.-%
of that particular gas, the error at 50 Vol.-% CO

2

and

50 Vol.-% N

2

is -0,135%.