Gas sampling and gas conditioning technology

3-4.20-ME

The sample gas flows through the pre-cooling stage 

and low-temperature stage . This cycle will

be changed automatically every three hours.
In the event of a brief (mains) power failure:

The current cycle status will be stored;

All LEDs go out;

The status alarm contact is activated;

The flow of sample gas is cut off externally, in case the sample gas pump or a solenoid valve in

the sample gas line are to be controlled by an alarm contact.


When the mains power is restored:

The controller starts up automatically;

The status alarm contact is deactivated;

The status alarm contact will release the flow of sample gas externally, in case the sample gas
pump or a solenoid valve in the sample gas line are controlled by an alarm contact.


In the event of a prolonged power failure, the controller will start up automatically like it happens for a
new start. Release is only done when the temperature is < -25°C.