Page 28: Temperature adjustment

background image

28

Gas sampling and gas conditioning technology

3-4.20-ME

18.2

TEMPERATURE SETTING FOR THE COOLER

The pre-cooling stage is set at the factory to a regulated temperature of +1°C.

Current-Temp.

R

i

> 100k

Setting-Temp.

P3

°C

+

-

Uref

Connect a d.c. voltmeter
to terminals 1, 2 and 3!

Uref

Utemp

R

C9

Figure 11 Temperature adjustment


Setting of the regulated temperature is done by trimming potentiometer P3, on the EC automatic
control board of the cooler. The setting range covers temperatures from 0°C to 20°C. Turning it to the
right sets a lower temperature, and turning it to the left sets a higher temperature.

Connecting an external voltmeter to plugs 2 and 3 allows the nominal set temperature to be read off
and controlled. A voltage value of 0.1V corresponds to a temperature of 1°C.

It is possible to measure and control the current temperature at measuring sockets 1 and 2.

N O T E !


Freezing up in the pre-cooling stage will endanger the operation of the
cooler unit. For this reason the cooler temperature must never fall below
+1°C!



19

TEMPERATURE SENSORS

The pre-cooling stage temperature sensor  is a PT100 element. STP 35 temperature sensors are
available for the low temperature stages  and .

19.1

CHECKING THE TEMPERATURE SENSOR OF THE PRE-COOLING STAGE

There are two methods for checking the PT100 element, as follows:

Voltage method


In order to check the sensor for the cooler currently in operation, the actual voltage at the measuring
sockets 1 and 2 of the EC automatic control board must be measured as per section 18.2 above.