beautypg.com

Closing down, Maintenance – M&C TechGroup EC-30_FD Operator's manual User Manual

Page 20

background image

20

Gas sampling and gas conditioning technology

3-4.20-ME

15

CLOSING DOWN

N O T E !


The location for the cooler must remain frost-free, even when the
unit has been switched off!


If the cooler unit is put out of operation for a short time no particular measures have to be taken.

We recommend purging the cooler with inert gas or ambient air, while the unit is put out of operation
for a longer time.

W A R N I N G !




Aggressive condensate is possible.
Wear protective glasses and proper protective clothing!



16

MAINTENANCE

The safety instructions specific to the plant and process are to be consulted prior to any maintenance
work!

W A R N I N G !



Dangerous voltage!
Before opening the housing please disconnect the cooler from the
mains supply!


The EC-30(/FD) cooler unit does not require any special maintenance intervals.

Depending on the degree of contamination of the ambient air the cooling fins of the condenser have
to be cleaned periodically with pressure air and the air suction filter units in the EC-F / EC-FD
universal unit have to be cleaned or replaced. They are located in a plug-in box underneath the
cooler unit. The filter elements are removed as follows:

Unclamp the plug-in box by turning the quick-lock screw plug 90



to the left;

Pull out the box;

Remove the crossbars holding the filter element by pulling with moderate force in the direction

of the filter element axis;

Remove and replace the filter elements.

Reassembly is done in the reverse order.