Pacing, Pacing -4, Nd “pacing” ch – Philips HEARTSTART XLT M3500B User Manual
Page 86

Pacing
8-4
Pacing
Pacing
To perform pacing:
1. Press
. The green LED next to
lights up and the dia-
logue box appears.
Figure 8-1 Pacing Display
The message
6WRS indicates that the pacing function is on but paced pulses
are not being delivered. The pacer turns on in the mode last used.
2. Verify that the dot markers appears near the middle of the QRS complexes
of the ECG.
If the dot markers do not appear, or are in the wrong location, adjust the
ECG size or select another lead. (See “Monitoring the ECG” in Chapter
4.)
The ECG will only appear if the 3- or 5-lead monitoring cable and elec-
trodes are used.
3. Press
to change to fixed mode, if R-wave detection is unreliable.
The pacing dialogue box displays the current mode. To switch back to
demand mode, press
again.
Pacer
Pacer
ENERGY
SELECT
HR ALARM
LEAD
SELECT
CHARGE
Shocks:0
00:00:8
78
Lead II
SpO2
150 J
Pacer Stop
Demand Mode
70ppm 30mA
Mode
Mode
- HB311 (54 pages)
- HB591 (32 pages)
- HB935 (56 pages)
- HP3621 (60 pages)
- HP3631 (108 pages)
- HB584 (96 pages)
- HB175 (76 pages)
- HB823 (72 pages)
- HB594 (88 pages)
- HB953 (36 pages)
- SCF290/15 (2 pages)
- 319 (84 pages)
- M3861A (102 pages)
- 862474 (182 pages)
- Twin Electronic SCF294/21 (2 pages)
- SCF156/00 (2 pages)
- SCF160 (2 pages)
- SCF302/01 (2 pages)
- Classic Breastfeeding Essentials 0 (2 pages)
- SCF294/02 (34 pages)
- AVENT SCF154/10 (2 pages)
- SCF155/06 (2 pages)
- Niplette SCF152/01 (2 pages)
- HeartStart MRx (2 pages)
- SCD241/01 (2 pages)
- M5066A (70 pages)
- SCF616/10 (2 pages)
- SCF300 (2 pages)
- SCF304/02 (34 pages)
- MICROLUX HC8900 (149 pages)
- Zinc-air ZA13B6A (2 pages)
- SCF304 (2 pages)
- AVENT SCF154/24 (2 pages)
- SCF156/01 (2 pages)
- AVENT SCF154/40 (2 pages)
- ISIS SCF292/15 (39 pages)
- AVENT SCF154/50 (2 pages)
- AVENT SCF213 (2 pages)
- ISIS iQ UNO (13 pages)
- ZA10B6A (2 pages)
- Avent SCF164 (2 pages)
- VIA Lids SCF614/10 (2 pages)
- SCF292/01 (2 pages)
- SCF290/13 (2 pages)