Defibrillating
6-4
Defibrillating in Manual Mode
Defibrillating
The defibrillation process is as follows:
1. Select Energy
The default energy setting is 150 joules. To change the energy setting,
press
repeatedly to cycle through the energy level choices. Stop
when the desired energy is displayed. The choices are 5, 10, 25, 50, 70,
100, 150, or 200 joules.
Figure 6-3 Energy Select Display
2. Charge
Press .
As the defibrillator charges, the current charge is displayed above the
shock counter. A charging tone beeps until the desired energy level is
reached, at which point you’ll hear a continuous charge tone.
(1(5*<
6(/(&7
ENERGY
SELECT
HR ALARM
LEAD
SELECT
CHARGE
Shocks:0
00:00:8
78
Pads
SpO2
150 J
&+$5*(
See also other documents in the category Philips Health and hygiene:
- HB311 (54 pages)
- HB591 (32 pages)
- HB935 (56 pages)
- HP3621 (60 pages)
- HP3631 (108 pages)
- HB584 (96 pages)
- HB175 (76 pages)
- HB823 (72 pages)
- HB594 (88 pages)
- HB953 (36 pages)
- SCF290/15 (2 pages)
- 319 (84 pages)
- M3861A (102 pages)
- 862474 (182 pages)
- Twin Electronic SCF294/21 (2 pages)
- SCF156/00 (2 pages)
- SCF160 (2 pages)
- SCF302/01 (2 pages)
- Classic Breastfeeding Essentials 0 (2 pages)
- SCF294/02 (34 pages)
- AVENT SCF154/10 (2 pages)
- SCF155/06 (2 pages)
- Niplette SCF152/01 (2 pages)
- HeartStart MRx (2 pages)
- SCD241/01 (2 pages)
- M5066A (70 pages)
- SCF616/10 (2 pages)
- SCF300 (2 pages)
- SCF304/02 (34 pages)
- MICROLUX HC8900 (149 pages)
- Zinc-air ZA13B6A (2 pages)
- SCF304 (2 pages)
- AVENT SCF154/24 (2 pages)
- SCF156/01 (2 pages)
- AVENT SCF154/40 (2 pages)
- ISIS SCF292/15 (39 pages)
- AVENT SCF154/50 (2 pages)
- AVENT SCF213 (2 pages)
- ISIS iQ UNO (13 pages)
- ZA10B6A (2 pages)
- Avent SCF164 (2 pages)
- VIA Lids SCF614/10 (2 pages)
- SCF292/01 (2 pages)
- SCF290/13 (2 pages)