Setting the Heart Rate Alarm
4-6
Monitoring the ECG
Setting the Heart Rate Alarm
The computed heart rate (number of detected QRS complexes per minute) is
displayed below the
sofkey, next to the
. The heart rate represents
the number of QRS complexes detected in a minute. A QRS beeper, if config-
ured to on, indicates each QRS complex detected.
If desired, a HR alarm may be set to alert you when the heart rate is outside
the specified limits. Limit choices are listed in Table 4-4.
To set a HR alarm, cycle through the limit choices by pressing
until
the desired limits are shown. The
then appears next to the heart rate value
to indicate that the HR alarm is set.
WARNING
Heart rate alarms are temporarily suspended in AED Mode during ECG analysis or
when
is pressed (for the duration of the paused period). Heart rate alarms are
also suspended while charging for defibrillation and delivering a shock.
WARNING
Heart rate displays and alarms function with internal and external pacemakers, but
they can be unreliable. Observe the patient closely if pacemakers are used.
Table 4-4 HR Alarm Limit Choices
Alarm If Over:
Or under:
100
30
140
60
160
90
200
120
+5 $/$50
+5 $/$50