Specifications

M3500B HeartStart XLT Defibrillator/Monitor

13-7

13

Noninvasive Pacing

Waveform: Monophasic Truncated Exponential.

Current Pulse Amplitude: 10 mA to 200 mA (5 mA resolution); accuracy 10

mA - 50 mA

5 mA, 50 mA - 200 mA

10%.

Pulse Width: 20 ms with accuracy +0, -5 ms.

Rate: 30 ppm to 180 ppm (10 ppm increments); accuracy

1.5%.

Modes: Demand or Fixed Rate.

Refractory Period: 340 msec (40 to 80 ppm); 240 msec (90 to 180 ppm).

SpO

2

/Pulse Oximetry

Accuracy with:

M1191A sensor - 1 standard deviation 70% to 100%,

2.5%.

M1192A sensor - 1 standard deviation 70% to 100%,

2.5%.

M1194A sensor - 1 standard deviation 70% to 100%,

4.0%.

NELLCOR sensors - 1 standard deviation 80% to 100%

3.0%.

Resolution: 1%.

SpO

2

Alarm Limits: Three preset low alarm limits: 90, 85, and 80%, default

is off.

INOP Alerts: Triggered by disconnected sensor, noisy signal, light interfer-
ence or low signal (non-pulsatile).

+