Philips HEARTSTART XLT M3500B User Manual
Page 60

Troubleshooting
4-8
Monitoring the ECG
Poor ECG signal
quality
z
The monitoring elec-
trodes are not making
proper contact with the
patient.
z
The monitoring elec-
trodes are outdated or
dried-out.
z
Radio frequency inter-
ference (RFI) is causing
artifact.
z
Check that the monitoring
electrodes are properly
applied. If necessary, pre-
pare the patient’s skin and
apply new electrodes.
z
Check the date code on
the electrodes. Do not
open the electrode pack-
age until immediately
prior to use.
z
Relocate or turn off equip-
ment that may be causing
RFI.
QRS beeper inau-
dible or beeps do
not occur with
each QRS com-
plex.
z
The QRS beeper is con-
figured to Off.
z
The amplitude of the
QRS complex is too
small to detect.
z
Check that the QRS
beeper is configured to
On.
z
Adjust the volume.
z
Adjust the size of the
ECG.
Table 4-5 Troubleshooting Tips (Continued)
Situation
Cause
Solution
- HB311 (54 pages)
- HB591 (32 pages)
- HB935 (56 pages)
- HP3621 (60 pages)
- HP3631 (108 pages)
- HB584 (96 pages)
- HB175 (76 pages)
- HB823 (72 pages)
- HB594 (88 pages)
- HB953 (36 pages)
- SCF290/15 (2 pages)
- 319 (84 pages)
- M3861A (102 pages)
- 862474 (182 pages)
- Twin Electronic SCF294/21 (2 pages)
- SCF156/00 (2 pages)
- SCF160 (2 pages)
- SCF302/01 (2 pages)
- Classic Breastfeeding Essentials 0 (2 pages)
- SCF294/02 (34 pages)
- AVENT SCF154/10 (2 pages)
- SCF155/06 (2 pages)
- Niplette SCF152/01 (2 pages)
- HeartStart MRx (2 pages)
- SCD241/01 (2 pages)
- M5066A (70 pages)
- SCF616/10 (2 pages)
- SCF300 (2 pages)
- SCF304/02 (34 pages)
- MICROLUX HC8900 (149 pages)
- Zinc-air ZA13B6A (2 pages)
- SCF304 (2 pages)
- AVENT SCF154/24 (2 pages)
- SCF156/01 (2 pages)
- AVENT SCF154/40 (2 pages)
- ISIS SCF292/15 (39 pages)
- AVENT SCF154/50 (2 pages)
- AVENT SCF213 (2 pages)
- ISIS iQ UNO (13 pages)
- ZA10B6A (2 pages)
- Avent SCF164 (2 pages)
- VIA Lids SCF614/10 (2 pages)
- SCF292/01 (2 pages)
- SCF290/13 (2 pages)