beautypg.com

7 performing synchronized cardioversion, Performing synchronized cardioversion, E “performing synchronized cardioversion – Philips HEARTSTART XLT M3500B User Manual

Page 79: 7performing synchronized cardioversion

background image

M3500B HeartStart XLT Defibrillator/Monitor

7-1

7

7

Performing Synchronized Cardioversion

Synchronized cardioversion is a Manual Mode function that allows you to
synchronize the defibrillator shock with the R-wave of the ECG being moni-
tored.

During synchronized cardioversion, the ECG being monitored is shown on
the display and may be derived from:

z

the multifunction defib electrode pads, or

z

the 3- or 5-lead monitoring electrodes.

When selecting a lead, choose the best lead that displays a large QRS com-
plex. The synchronized shock is delivered through the mulifunction defib
electrode pads, regardless of the lead being monitored.

This chapter describes how to perform synchronized cardioversion with the
HeartStart XLT.

NOTE

See Chapter 4, “Monitoring the ECG” for information on how to apply elec-
trodes and select a lead.