Operational Checks

M3500B HeartStart XLT Defibrillator/Monitor

11-3

11

Every Shift

Perform a “Shift/System Check” every shift (see “Shift/System Check” on
page 11-4) to verify that the HeartStart XLT is functioning properly and to
ensure that necessary supplies and accessories are present and ready for use.

Every Month

Check expiration dates on multifunction defib electrode pads and monitoring
electrodes every month. Replace them if the expiration date has passed.

Every Three Months

Perform a "Battery Capacity Test" on each battery, every three months, to
ensure that your batteries meet the specifications for safe and effective use.