Operational Checks
M3500B HeartStart XLT Defibrillator/Monitor
11-5
11
The report also lists additional checks that you should do. Perform each of
these checks and record the results. The guidelines for completing the checks
are as follows:
Defibrillator Inspection - make sure the HeartStart XLT is clean, clear of
objects on top and has no visible signs of damage.
Cables/Connectors - make sure there are no cracks, broken wires, or other
visible signs of damage. Make sure the connectors engage securely.
Supplies - make sure the carrying case has:
z
two sets of multifunction defib electrode pads in sealed packages,
within the expiration date
z
an adequate supply of monitoring electrodes, within the expiration
date
z
alcohol wipes
z
hand towel
z
scissors
z
a razor
z
an extra roll of printer paper
z
a spare charged battery
z
a Data Card
z
SpO
2
sensors (if monitoring SpO
2
)
- HB311 (54 pages)
- HB591 (32 pages)
- HB935 (56 pages)
- HP3621 (60 pages)
- HP3631 (108 pages)
- HB584 (96 pages)
- HB175 (76 pages)
- HB823 (72 pages)
- HB594 (88 pages)
- HB953 (36 pages)
- SCF290/15 (2 pages)
- 319 (84 pages)
- M3861A (102 pages)
- 862474 (182 pages)
- Twin Electronic SCF294/21 (2 pages)
- SCF156/00 (2 pages)
- SCF160 (2 pages)
- SCF302/01 (2 pages)
- Classic Breastfeeding Essentials 0 (2 pages)
- SCF294/02 (34 pages)
- AVENT SCF154/10 (2 pages)
- SCF155/06 (2 pages)
- Niplette SCF152/01 (2 pages)
- HeartStart MRx (2 pages)
- SCD241/01 (2 pages)
- M5066A (70 pages)
- SCF616/10 (2 pages)
- SCF300 (2 pages)
- SCF304/02 (34 pages)
- MICROLUX HC8900 (149 pages)
- Zinc-air ZA13B6A (2 pages)
- SCF304 (2 pages)
- AVENT SCF154/24 (2 pages)
- SCF156/01 (2 pages)
- AVENT SCF154/40 (2 pages)
- ISIS SCF292/15 (39 pages)
- AVENT SCF154/50 (2 pages)
- AVENT SCF213 (2 pages)
- ISIS iQ UNO (13 pages)
- ZA10B6A (2 pages)
- Avent SCF164 (2 pages)
- VIA Lids SCF614/10 (2 pages)
- SCF292/01 (2 pages)
- SCF290/13 (2 pages)