ERC Protocol
M3500B HeartStart XLT Defibrillator/Monitor
3-17
3
If CPR is needed, press
. While paused, a timer indicates the elapsed
time and the total duration of the Pause state, as shown:
Figure 3-15
The total pause duration depends on the event preceding the Pause state. If
you entered the Pause state:
z
at the completion of a shock series or shortly after a shock is deliv-
ered, the duration is equal to the
3RVW 6KRFN &35 7LPHU configuration set-
ting (the default setting is 60 seconds).
z
when no shock was advised, the duration is equal to the
16$ 7LPHU
configuration setting, where NSA is an acronym for No Shock
Advised (the default setting is 180 seconds).
3$86(
Shocks: 3
00:01:40
Paused
RESUME
ANALYZE
Timer: 35 / 60
Elapsed Time
Total Pause Duration
See also other documents in the category Philips Health and hygiene:
- HB311 (54 pages)
- HB591 (32 pages)
- HB935 (56 pages)
- HP3621 (60 pages)
- HP3631 (108 pages)
- HB584 (96 pages)
- HB175 (76 pages)
- HB823 (72 pages)
- HB594 (88 pages)
- HB953 (36 pages)
- SCF290/15 (2 pages)
- 319 (84 pages)
- M3861A (102 pages)
- 862474 (182 pages)
- Twin Electronic SCF294/21 (2 pages)
- SCF156/00 (2 pages)
- SCF160 (2 pages)
- SCF302/01 (2 pages)
- Classic Breastfeeding Essentials 0 (2 pages)
- SCF294/02 (34 pages)
- AVENT SCF154/10 (2 pages)
- SCF155/06 (2 pages)
- Niplette SCF152/01 (2 pages)
- HeartStart MRx (2 pages)
- SCD241/01 (2 pages)
- M5066A (70 pages)
- SCF616/10 (2 pages)
- SCF300 (2 pages)
- SCF304/02 (34 pages)
- MICROLUX HC8900 (149 pages)
- Zinc-air ZA13B6A (2 pages)
- SCF304 (2 pages)
- AVENT SCF154/24 (2 pages)
- SCF156/01 (2 pages)
- AVENT SCF154/40 (2 pages)
- ISIS SCF292/15 (39 pages)
- AVENT SCF154/50 (2 pages)
- AVENT SCF213 (2 pages)
- ISIS iQ UNO (13 pages)
- ZA10B6A (2 pages)
- Avent SCF164 (2 pages)
- VIA Lids SCF614/10 (2 pages)
- SCF292/01 (2 pages)
- SCF290/13 (2 pages)