ERC Protocol

M3500B HeartStart XLT Defibrillator/Monitor

3-17

3

If CPR is needed, press

. While paused, a timer indicates the elapsed

time and the total duration of the Pause state, as shown:

Figure 3-15

The total pause duration depends on the event preceding the Pause state. If
you entered the Pause state:

z

at the completion of a shock series or shortly after a shock is deliv-
ered, the duration is equal to the

3RVW 6KRFN &35 7LPHU configuration set-

ting (the default setting is 60 seconds).

z

when no shock was advised, the duration is equal to the

16$ 7LPHU

configuration setting, where NSA is an acronym for No Shock
Advised (the default setting is 180 seconds).

3$86(

Shocks: 3

00:01:40

Paused

RESUME

ANALYZE

Timer: 35 / 60

Elapsed Time
Total Pause Duration