Configuring the HeartStart XLT

M3500B HeartStart XLT Defibrillator/Monitor

10-11

10

Table 10-2 AED Settings:

Parameter

Description

Setting

Choices

$(' 6KRFN 6HULHV

Defines the maximum number of shocks to
deliver before prompting

&KHFN 3DWLHQW &KHFN

3XOVH ,I 1HHGHG %HJLQ &35.

2, 3, 4

6KRFN 6HULHV 7LPHU Defines the number of seconds that must pass

before the next shock is considered the first
shock of a new shock series, rather than the
next shock of the current shock series.

30, 60, 90,
120, 150, 180,
210, Off

'HYLFH ,QLWLDWHG

$QDO\VLV

Initiates ECG analysis when the HeartStart
XLT is turned on in AED Mode.

On, Off

$XWRPDWLF

5H$QDO\VLV

Initiates ECG analysis in between shocks
within a shock series.

On, Off

5K\WKP 0RQLWRULQJ

Monitors the ECG for potentially shockable
rhythms when the HeartStart XLT is not ana-
lyzing, defibrillating, or paused.

On, Off

´&KHFN 3DWLHQWµ

7LPHU

Defines how often (in seconds) the

&KHFN

3DWLHQW prompt is repeated when Rhythm Mon-
itoring detects a potentially shockable rhythm.

30, 45, 60, 90,
Off

(XURSHDQ 3URWRFRO

Modifies Pause state prompts and replaces the
Pause Timer with either the Post Shock CPR
Timer or the "NSA" Timer, depending on the
event preceding the Pause state.

Off/On

If

(XURSHDQ SURWRFRO is set to On, the following two configuration choices appear:

3RVW 6KRFN &35

7LPHU

Defines the duration of the Pause time (in sec-

onds) when

is pressed and the time

since the last shock is less than or equal to the
6KRFN 6HULHV 7LPHU setting - typically at the com-
pletion of a shock series.

30, 60, 120,
180

3$86(