beautypg.com

Procom VOR/ILS Analyzer EVS200 User Manual

Page 60

background image

VOR/ILS Analyzer EVS200

0796.1800.02

E-8

51

Chapter 3: Operation

Command AM7

Kommando

Funktion

AM6

query of AM-modulation depth ID

The command AM7 reads the actual AM-modulation depth (ID) in the BEACON mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

AM7
am7

e.g.

10.2%

e.g.

10.2%

Command AF4

command

function

AF4

query of AF-frequency 3000 Hz

The command AF4 reads the actual AF frequency (F1) in the BEACON mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

AF4
af4

e.g.

3000.0Hz

e.g.

3000.0Hz

Command AF5

command

function

AF5

query of AF-frequency 1300 Hz

The command AF5 reads the actual AF frequency (F2) in the BEACON mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

AF5
af5

e.g.

1300.0Hz

e.g.

1300.0Hz