Chapter 3: Operation

Unit related control commands

Command RF (frequency in kHz)

command

function

RF (frequency in kHz) setting or scanning receiving frequency

With the command RF (frequency in kHz) a frequency input is possible in the ranges of 107000 to
119000 kHz and 319000 to 341000 kHz. With this command it is also possible to scan the tuned
frequency when the command RF is transmitted to the EVS200 without additionally parameters. The
unit responses with the frequency information.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

input:

RF107000
rf107000

READY
READY

query:

RF
rf

107000kHz
107000kHz

Command LO

command

function

LO

set unit local mode

Because local operation is automatically locked when the unit is controlled by a controller, the com-
mand LO (local) performs further local operation at the unit.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

input:

LO
lo

READY
READY

Command EC0

command

function

EC0

selection of communication

EVS200 does not retransmit the received control string.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

EC0
ec0

READY
READY