VOR/ILS Analyzer EVS200

0796.1800.02

E-8

36

Chapter 3: Operation

Command DD1

command

function

DD1

query of the DDM value in

µµµµ

A

The command DD1 reads the actual DDM value in µA in the ILS mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

DD1
dd1

e.g.

194.32uA

e.g.

194.32uA

Command SD0

command

function

SD0

SDM value query

The command SD0 reads the actual SDM value (value without dimension) in the ILS mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

SD0
sd0

e.g.

0.8006

e.g.

0.8006

Command SD1

Command

function

SD1

query of SDM value in

µµµµ

A

The command SD1 reads the actual DDM value in µA in the ILS mode.

Example:

command from the controller

response of the VOR/ILS Analyzer EVS200

query:

SD1
sd1

e.g.

774.87uA

e.g.

774.87uA