P907 Programmer Interface

Press ENTER

to activate the programmer interface (two way infrared

communications) test.
On successful test completion, PASS is displayed. Otherwise, FAIL is displayed.

P910 Relay

Access this parameter to display the current state of the specified relay.

Press ENTER

to alter the relay state (energized / de-energized) or enter the value

corresponding to the specific relay state desired.

P911 mA Output Value

Access this parameter to display the current value of the mA output.

Additionally, this feature may be used to enter a desired value. The mA output
immediately assumes the value entered regardless of any restrictions programmed.

P912 Transducer Temperature

Use this feature to display the temperature in °C (as monitored by the connected

transducer).

Err is displayed if the transducer is not equipped with an internal temperature sensor.

P913 Sensor Temperature

Access this parameter to display the temperature in °C (as monitored by the TS-3).

OPEn is displayed if a TS-3 is not connected.

Values

0 = de-energized
1 = energized

Primary Index

mA output

Values

Range: 0.00 to 25.00

Related

P200 mA Output Range

P201 mA Output Function

Values

Range: -50 to 150

Values

Range: -50 to 150