beautypg.com

P651 sound velocity calibration – Siemens sitrans LU01 User Manual

Page 61

background image

7ML19985BE02

SITRANS LU01 – INSTRUCTION MANUAL

Page 55

m

mmmm

Pa
ra

m

e

te

rs

Before using this feature, verify the following parameters are correct:

Empty (P006)

Temperature (P664)

Offset Reading (P062)

Offset Calibration

Begin with a steady level.

1.

Press TRANSDUCER

to display the calculated reading.

2.

Repeat Step One at least five times to verify repeatability.

3.

Measure the actual reading (use tape measure).

4.

Enter the actual value.

The deviation between the entered Empty (P006) value and the calibrated Empty value is
stored in Offset Correction (P652).

P651 Sound Velocity Calibration

Changes the speed of sound constant.

Condition for use of this feature

The acoustic beam atmosphere is other than air

The acoustic beam atmosphere temperature is unknown

The Reading accuracy is acceptable at higher material levels only

For best results, calibrate with the level at a known value near empty.

Using Sound Velocity Calibration

Ensure a steady level at some low value (P653 and P654 adjusted accordingly)

1.

Allow sufficient time for the vapor concentration to stabilize.

2.

Press TRANSDUCER

to display the calculated reading.

3.

Repeat Step Two at least five times to verify repeatability.

4.

Measure the actual reading (e.g. with a tape measure).

5.

Enter the actual value.

Repeat this procedure if the atmosphere type, concentration, or temperature conditions
are different from when the last sound velocity calibration was performed.

Values

Range: -999 to 9999

Related

P653 Velocity

P654 Velocity at 20°C

Note:

In gasses other than air, the temperature variation may not correspond

with the speed of sound variation. Turn off temperature sensor and use a fixed
temperature.