beautypg.com

Transducer, Options, Cable – Siemens sitrans LU01 User Manual

Page 13: Transducer options cable

background image

7ML19985BE02

SITRANS LU01 – INSTRUCTION MANUAL

Page 7

mm

mmm

Spec

if

icat

ions

Transducer

Compatible Models

• STH and Echomax

®

series

• Refer to the associated instruction manual.

Options

Temperature Sensor

• TS-3

SmartLinx Modules

• Supported protocols: PROFIBUS DP

Allen-Bradley®

1

Remote I/O

Modbus® RTU
DeviceNet

TM

Dolphin Plus

• Windows

®

-compatible configuration software connected to the unit via infrared

Comverter link
Refer to associated product documentation.

Cable

Transducer

• RG-62 A/U (or equivalent), 365 m (1,200 ft) max.

See transducer instructions for short extensions (in grounded metal conduit, separate
from other wiring)

mA Output

• Belden 8760, shielded / twisted pair, 18 AWG (0.75 mm

2

) or equivalent or Belden 9552,

shielded / two twisted pair, 18 AWG (0.75 mm2)

• maximum separation 1,500 m (5,000 ft)

Synchronisation

• Belden 8760, shielded / twisted pair, 18 AWG (0.75 mm

2

) or equivalent

Relays

• No shielded cable necessary

Temperature sensor

• Belden 8760 shielded / twisted pair, 18 AWG (0.75 mm

2

) or equivalent

• 365 m (1,200 ft) per TS-3 maximum

RJ11 Link

• No shielded cable necessary
• maximum length 3 m (10 ft)

1.

Allen-Bradley is a registered trademark of Rockwell Automation. Modbus is a
registered trademark of Schneider Electric. DeviceNet is a trademark of ODVA
(Open DeviceNet Vendor Association).