SITRANS LU01 – INSTRUCTION MANUAL

Page 5

mm

mmm

Spec

if

icat

ions

Specifications

Electronics

Power

• AC model:

100/115/200/230 V AC ±15%, 50/60 Hz, 31 VA

• DC model:

18 to 30 V DC, 25 W

Environmental

• location

indoor / outdoor

• altitude

2000 m max

• ambient temperature –20 to 50 °C (–5 to 122 °F)
• relative humidity

suitable for outdoor (Type 4X / NEMA 4X / IP65 enclosure)

• installation category II
• pollution degree:

4

Scan Points

• 1 point per SITRANS LU01 max.
• frequency independent

Range

• Level Measurement: 0.3 m (1 ft) to 60 m (200 ft) max.

Accuracy

• 0.25% of range or 6 mm (0.24"), whichever is greater

Resolution

• 0.1% of program range

1

or 2 mm (0.08"), whichever is greater

Memory

• EEPROM (non-volatile) no back-up battery required

Programming

• via removable programmer or optional Dolphin Plus software

Display

• custom graphics backlit LCD with 51 mm (2 in) x 127 mm (5 in) viewing area

Synchronization:

• up to 16 SITRANS LU01 units can be synchronized together

Temperature Compensation

• –50 to 150 °C (–58 to 302 °F)
• integral temperature sensor in transducer
• TS-3 temperature sensor
• programmable fixed temperature

Temperature Error

• with compensation:

0.09% of range

• fixed temperature:

0.17% / °C deviation from programmed temperature.

1.

program range is defined as the empty distance to the face of the transducer

(P006) plus any range extension (P801).