P843 long shot frequency, P844 short shot width, P845 long shot width – Siemens sitrans LU01 User Manual
Page 85: P850 short shot bias
7ML19985BE02
SITRANS LU01 – INSTRUCTION MANUAL
Page 79
m
mmmm
Pa
ra
m
e
te
rs
P843 Long Shot Frequency
Adjust the long shot transmit pulse frequency (in kHz).
This feature is automatically altered when Transducer (P004) is altered.
P844 Short Shot Width
Adjust the width (in ms) of the short shot transmit pulse.
This feature is automatically altered when Transducer (P004) is altered.
P845 Long Shot Width
Adjust the width (in ms) of the long shot transmit pulse.
This feature is automatically altered when Transducer (P004) is altered.
P850 Short Shot Bias
Use this feature to slant the echo evaluation in favour of the short shot echo when both
short and long shots are evaluated (see Shot Mode, P803).
Values
Range: 10 to 60 kHz
Altered By
•
P004 Transducer
Related
•
P841 Long Shot Number
•
P842 Short Shot Frequency
•
P844 Short Shot Width
•
P845 Long Shot Width
Values
Range: 0.000 to 5.000
Altered By
•
P004 Transducer
Related
•
P840 Short Shot Number
•
P842 Short Shot Frequency
•
P845 Long Shot Width
•
P850 Short Shot Bias
•
P851 Short Shot Floor
•
P852 Short Shot Range
Values
Range: 0.000 to 5.000
Altered By
•
P004 Transducer
Related
•
P841 Long Shot Number
•
P844 Short Shot Width
•
P843 Long Shot Frequency
Values
Range: 0 to 100
Preset: 20
Related
•
P803 Shot / Pulse Mode
•
P840 Short Shot Number
•
P842 Short Shot Frequency
•
P844 Short Shot Width
•
P851 Short Shot Floor
•
P852 Short Shot Range