beautypg.com

Ma output failsafe (p219), Standard data logging (p300 to p321), Record temperatures (p300 to p303) – Siemens sitrans LU01 User Manual

Page 56: P219 ma output failsafe, P300 temperature, transducer maximum

background image

Page 50

SITRANS LU01 – INSTRUCTION MANUAL

7ML19985BE02

Pa

ra

m

e

te

rs

mA Output Failsafe (P219)

P219 mA Output Failsafe

Use for failsafe operation, independent of the Failsafe Material Level (P071).

Selecting an independent mA Failsafe option:

1.

Press MODE

to display the Auxiliary Function symbol.

2.

Press ARROWS

to scroll access the failsafe options.

3.

Press ENTER

when the desired option displayed.

Or, to produce an mA output at a specific value, enter the value required. This is used only
if mA output is allocated to a transducer (P201 = 1 to 4).

Standard Data Logging (P300 to P321)

All records can be reset by pressing the CLEAR

keys.

Record Temperatures (P300 to P303)

These features display the high and/or low temperatures in °C. When a parameter
relating to a TS-3 Temperature Sensor is accessed, the Index Type display changes to the
TS-3 symbol .

If the unit is powered up without a temperature sensor connected, the value –50 °C is
displayed. This information can help trace problems with both built in and external
temperature sensors.

P300 Temperature, Transducer Maximum

Shows the highest temperature encountered, as measured by the temperature sensor in

the transducer (if applicable).

Press CLEAR

keys to reset the log after a short circuit on the transducer wiring.

Primary Index

mA output

Values

Range: 0.000 to 22.00
OFF

*

mA output responds to Failsafe Material Level (P071).

HOLd

last known value is held until normal operation
resumes

LO

produce the Empty mA output immediately

HI

produce the Span mA output immediately

Related

P201 mA Output Function

Values

Range: - 50 to 150°C (view only)
Preset: - 50°C