beautypg.com

P702 filling indicator, P703 emptying indicator, P704 rate filter – Siemens sitrans LU01 User Manual

Page 65

background image

7ML19985BE02

SITRANS LU01 – INSTRUCTION MANUAL

Page 59

m

mmmm

Pa
ra

m

e

te

rs

P702 Filling Indicator

The fill rate required to activate the LCD Filling indicator ( ).

This value (in Units (P005) or % of Span (P007) per minute) is automatically set to 1/10 of
the Max Fill Rate (P700).

P703 Emptying Indicator

The empty rate required to activate the LCD Emptying indicator ( ).

This value (in Units (P005) or % of Span (P007) per minute) is automatically set to 1/10 of
the Max Empty Rate (P701).

P704 Rate Filter

Damps Rate Value (P707) fluctuations.

Enter the time or distance interval over which the Rate Value is to be calculated before
the display updates.

This is automatically altered along with Maximum Process Speed (P003).

This value automatically alters the Rate Update Time (P705) and / or Rate Update
Distance (P706). Alternatively, these parameter values may be altered independently.

Values

Range: -999 to 9999

Altered by

P003 Maximum Process Speed

Related

P005 Units

P007 Span

P700 Max Fill Rate

Values

Range: -999 to 9999

Altered by

P003 Maximum Process Speed

Related

P005 Units

P007 Span

P701 Max Empty Rate

Values

0

Rate display not required

Filtered Output
1

Continuously filtered and updated

Interval Output
2

1 minute or 50 mm (2 in)

3

5 minutes or 100 mm (3.9 in)

4

10 minutes or 300 mm (11.8 in)

5

10 minutes or 1000 mm (39.4 in)

Alters

P707 Rate Value

Altered by

P003 Maximum Process Speed

Related

P705 Rate Update Time/ P706 Rate Update Distance