Test (p900 to p913), P851 short shot floor, P852 short shot range – Siemens sitrans LU01 User Manual
Page 86: P900 software revision number
Page 80
SITRANS LU01 – INSTRUCTION MANUAL
7ML19985BE02
Pa
ra
m
e
te
rs
P851 Short Shot Floor
Enter the minimum echo strength (in dB above 1 μV rms) derived from a short shot to be
considered for evaluation.
P852 Short Shot Range
Enter the maximum distance in Units (P005) to be measured using short shot echoes.
This feature is automatically altered when Transducer (P004) is altered.
Test (P900 to P913)
P900 Software Revision Number
View the EPROM Rev. #.
P901 Memory
Press ENTER
to activate the SITRANS LU01 memory test.
Values
Range: 30 to 100
Preset: 50
Related
•
P840 Short Shot Number
•
P842 Short Shot Frequency
•
P844 Short Shot Width
•
P850 Short Shot Bias
•
P852 Short Shot Range
Values
Range: 0.000 to 9999
Altered By
•
P004 Transducer
Related
•
P840 Short Shot Number
•
P842 Short Shot Frequency
•
P844 Short Shot Width
•
P850 Short Shot Bias
•
P851 Short Shot Floor
Note:
These parameters are for Siemens Milltronics service personnel only.
Values
Range: 00.00 to 99.99 (view only)
Values
Display: view only
PASS
(memory test successful)
F1
RAM
F2
NOVRAM
F3
EEPROM
F4
EPROM