beautypg.com

Mt-01 hold message – Siemens sitrans LU01 User Manual

Page 102

background image

Page 96

SITRANS LU01 – INSTRUCTION MANUAL

7ML19985BE02

mmmm

m

Te

c

h

n

ical Ref

e

re

nce

Status B
1 character, $0 to $F (convert to binary, e.g. scan point is in high alarm)

Status C
1 character, $0 to $F (convert to binary, e.g. scan point is not in rate or band alarm)

MT-01 Hold Message

Field Name

Definition

Temp

2 characters, $32 to $FA (convert to decimal and subtract 100, e.g. tempera-
ture = 20 °C)

Percent

5 characters, 00.00 to 9999. (e.g. current level = 85.00% of span, EEEE. = field
overflow)

End Of Message

CR ($0D)

Field Name

Definition

Start Of Message STX ($02)

Message Type

2 characters, 01 (indicates the SITRANS LU01 was removed from the
RUN mode)

End Of Message

CR ($0D)

# # # #

1 = low low alarm

1 = low alarm
1 = high alarm
1 = high high alarm

MSB

LSB

# # # #

1 = rate alarm 2

1 = rate alarm 1

1 = band alarm 1
1 = band alarm 2

MSB

LSB

start of message

message type

end of message