SITRANS LU01 – INSTRUCTION MANUAL

7ML19985BE02

Pa

ra

m

e

te

rs

To manually record a profile

Press TRANSDUCER

to fire the transducer and record the echo profile into the

internal scope buffer for display.

For differential or average operation (P001 = 4 or 5), access Scope Displays (P810)
parameter to select the transducer number.

To save a manual record

Press ENTER

to copy the echo profile record in the scope buffer and save it in the

selected address in the record library. The parameter value field displays the new record
information.

To display a record

Press

key to enter display auxiliary mode.

Then press TRANSDUCER

to copy the current echo profile into the scope buffer for

display on an oscilloscope or Dolphin Plus

To delete a record

Press CLEAR

and then ENTER

to delete the echo profile record in the selected

address. The value returns to - - - -.

P331 Auto Record Enable

Use to enable/disable the Auto Profile Record function.

P333 Auto Record Interval

Enter the time to elapse after an Auto Profile Record is saved before another Auto Profile

Record can be saved (subject to all other restrictions).

Auto Record ON and OFF Setpoints (P334 to P337)

Use Auto Record ON Setpoint (P334) and Auto Record OFF Setpoint (P335) to define the
boundaries within which the level must be, for the resultant Echo Profile to be considered
for an Auto Profile Record.

If ---- is displayed for either P334 or P335, Auto Profile Records are saved regardless of
current level (subject to all other restrictions).
Enter the level value in Units (P005) or percent of Span (P007) as referenced to
Empty (P006).

Values

Range: 0 to 1
0

*

Off

1

On

Values

Range: 0.0 to 9999 (minutes)
Preset: 120