HP OMNIBER 718 HP 37718A User Manual
Page 72
62
Selecting Test Features
Using Pointer Graph Test Function
If, during a long term measurement (4 days), an event occurs at a
particular time each day, a short term measurement can be made at
the identified time to gain more detail of the event.
3
Choose the POINTER UNDER TEST type.
4
Press
to start the measurement.
TIP:
If the event occurs outside normal working hours, a Timed Start
measurement can be made.
1 SEC - display window of approximately 5 minutes.
5 SECS - display window of approximately 25 minutes.
20 SECS - display window of approximately 1 hour 40 minutes.
1 MIN - display window of approximately 5 hours.
5 MIN - display window of approximately 1 day.
20 MIN - display window of approximately 4 days.
RUN/STOP