beautypg.com

Start the jitter transfer measurement – HP OMNIBER 718 HP 37718A User Manual

Page 105

background image

95

Making Measurements

Measuring Jitter Transfer

HOW TO:

Start the Jitter Transfer Measurement

N OT E

The Jitter Transfer measurement must be started within 10 minutes of
the completion of calibration.

1

After the CALIBRATION is completed, remove the back to back
connection from the DSn or SONET or optical interfaces.
If the measurement is to be made at a network equipment monitor
point, choose

on the DSn or SONET

display

before making the jitter transfer measurement.

2

Choose MODE

on the

display and

press

.

The measurement’s progress can be monitored on the
display. At the end of the test the graph can be viewed on the

display.

MONITOR

RECEIVE

MEASURE

TRANSMIT

JITTER

RUN/STOP

TRANSMIT

RESULTS

JITTER