Inserting an external dsn payload/test signal, Insert 34 mb/s & ds3, Inserting an external dsn payload/test signal 43 – HP OMNIBER 718 HP 37718A User Manual
Page 53: Hosen, see “inserting an external dsn payload/test
43
Selecting Test Features
Inserting an External DSn Payload/Test Signal
Inserting an External DSn Payload/Test Signal
Description
You can insert a DSn signal from external equipment into the SONET
signal, or you can insert 2 Mb/s or DS1 into the structured DSn signal, as
shown in the table below. DS3 and 34 Mb/s can only be inserted if
SONET is chosen as the receive interface. 2 Mb/s or DS1 can be inserted
from a structured or non-structured SONET payload and from a
structured DSn signal.
HOW TO:
Insert 34 Mb/s & DS3
1
Connect the external payload to the 75
Ω
IN port of the PDH/DSn
receive module.
2
Set up the required transmit SONET interface, and choose VT
PAYLOAD
or
as required.
RATE
Availability
Option
DS3
SONET
011 Only
34Mb/s
SONET
010 and 011
2Mb/s
DSn & SONET
010 and 011
DS1
DSn & SONET
011 Only
INSERT 34 Mb/s
INSERT DS3