Rate:current:ac, Rate:resistance – RIGOL DM3058 User Manual
Page 73

RIGOL
Programming Guide for DM3058/DM3058E
3-48
4. :RATE:CURRent:AC
Command Format:
:RATE:CURRent:AC?
:RATE:CURRent:AC {F|M|S}
Function:
The commands query and set the rate of AC current measurement.
Explanations:
:RATE:CURRent:AC? - Queries the current measuring rate of AC current. The
command is valid only when AC current function is enabled.
:RATE:CURRent:AC {F|M|S} - Sets the desired measuring rate of AC current. For
the range and the explanation of each parameter, see Table 3-10.
Return Format:
The query returns the parameter related to current rate (see Table 3-10).
Example:
Sets the rate under AC current measurement to Medium.
:RATE:CURRent:AC M
5. :RATE:RESistance
Command Format:
:RATE:RESistance?
:RATE:RESistance {F|M|S}
Function:
The commands query and set the rate of 2-wire resistance measurement.
Explanations:
:RATE:RESistance? - Queries the current measuring rate of 2-wire resistance. The
command is valid only when 2-wire resistance function is enabled.
:RATE:RESistance {F|M|S} - Sets the desired measuring rate of 2-wire resistance.
For the range and the explanation of each parameter, see Table 3-10.
Return Format: