Measure:current:ac:range, Measure:resistance – RIGOL DM3058 User Manual
Page 60

RIGOL
Programming Guide for DM3058/DM3058E
3-35
Discrete
{0|1|2|3}
1
Explanations:
:MEASure:CURRent:AC
The measurement type will change into “Manual” automatically when you use this
command.
For the parameter range, see table below.
Table 3-7 AC current range
Parameter
Range
0
20 mA
1
200 mA
2
2 A
3
10 A
MIN
20 mA
MAX
10 A
DEF
200 mA
Example:
Sets the AC current range to the maximum (10A).
:MEASure:CURRent:AC MAX or :MEASure:CURRent:AC 3
17. :MEASure:CURRent:AC:RANGe?
Command Format:
:MEASure:CURRent:AC:RANGe?
Function:
The command queries the AC current range.
Explanations:
The command is valid only when AC current measurement is enabled.
Return Format:
The command returns: 0, 1, 2, 3 or 4 (see Table 3-7).
18. :MEASure:RESistance?