Sense:]detector:bandwidth – RIGOL DM3058 User Manual
Page 142

RIGOL
Programming Guide for DM3058/DM3058E
4-49
Function:
The commands set and query the aperture time of period measurement.
Explanations:
[SENSe:]PERiod:APERture {0.01|0.1|1|MINimum|MAXimum} – Sets the desired
aperture time for period measurement. The unit is s, MIN=10ms, MAX=1s, and
the default is 100ms.
[SENSe:]PERiod:APERture? [MINimum|MAXimum] – Queries the aperture time of
period measurement.
Return Format:
The query returns the current aperture time of period measurement.
30. [SENSe:]DETector:BANDwidth
Command Format:
[SENSe:]DETector:BANDwidth {3|20|200|MINimum|MAXimum}
[SENSe:]DETector:BANDwidth? [MINimum|MAXimum]
Function:
The commands set and query the possible minimum frequency from signals
inputted. The multimeter will select a measurement rate according to your
frequency setting such as S, M (default) or F. Note that the command is only
responsed without practical operation.
Explanations:
[SENSe:]DETector:BANDwidth {3|20|200|MINimum|MAXimum} – Sets the
desired minimum frequency for signals inputted. The unit is Hz, MIN=3Hz,
MAX=200Hz.
[SENSe:]DETector:BANDwidth? [MINimum|MAXimum]? – Queries the minimum
frequency from signals inputted.
Return Format:
The query returns “3”, “20” or “200”.